The Swedish NMR Centre is Centre is a national facility for metabolomics by NMR within the framework of SwedNMR and SciLifeLab.
Metabolomics – what is it?
Metabolomics research focuses on the analysis of small (<1.5kDa) molecules (metabolites) and their temporal and systematic variation altered by internal or external factors. In short, metabolomics is the identification and quantification of all metabolites in a biological sample detected by the present technology at a given condition and/or point in time. For NMR, this encompasses about 50 to a few hundred metabolites depending on the matrix analyzed. At the Swedish NMR Centre we have experience of a number of different biological matrices for metabolomics analysis, the majority being human origin (serum/plasma, urine, feces and extracts of different tissues such as muscle, liver and brain) but also including samples from model organisms such as mouse, rat, yeast, insects, macroalgae, mammalian cell cultures and bacteria.
The two dominant analytical techniques in metabolomics is mass spectrometry and NMR. The former is available to researchers in Sweden primarily through the Swedish Metabolomics Centre, which also has a node in Gothenburg in the Chalmers Mass Spectrometry Infrastructure. The Swedish NMR Centre in Gothenburg and Umeå offers NMR-based metabolomics, with access to both expertise in sample collection for solid-state NMR (e.g. biopsies) and for bioliquids, and also support for statistical analysis of collected data is available.
In metabolomics, the preanalytical phase of planning, collecting samples and sample preparation is of utmost importance in order to avoid unwanted confounding variance in the resulting spectroscopic data. For a successful project, it is necessary to consider even minute practical considerations:
Is there a sampling SOP in place, e.g. stating what sampling tube, timing, centrifugation speed, temperature storage, sample transport method etc. should be used?
If the sampling occurs at different sites, by different people, has any effort in unfying the SOP been done?
A central mantra in planning for a metabolomics project should be 'same, same, same' to ensure that, even though maybe a suboptimal sampling scheme was used, at least a comparison between samples that have had a similar history before ending up in the NMR spectrometer will be sensible. Case study
The project is about the effect of gastric bypass surgery and related medication on diabetic kidney disease (DKD). The samples were urine that were prepared according to the Bruker IVDr SOP for human urine samples, i.e. dilution 9:1 sample:buffer before transfer to SampleJet NMR tube racks. The sample preparation was done with a Bruker SamplePro liquid handling robot (nickname "Sebulba") to minimize human handling error. All sample were kept cool throughout the preparation process and also in the sample changer before going into the spectrometer. Acquisition order was randomized to minimize batch effects.
If possible, repeated measures or matched pairs, e.g. the same participants take part in each condition of the independent variables, and the participants in different groups share characteristics as age and gender, is preferred. This also reduces necessary sample sizes.
When e.g. larger differences between two groups can be expected, tens of samples might be sufficient, but realistic sizes are more often in the order of hundreds of samples.
Specific for metabolomics is that the result often depends on the pre-analytical treatment, e.g. sample tube type, time to centrifugation, freeze cycles, etc. It is important that control samples have been treated in the same way as the 'real' samples in a study.
SNC GU has experience in acquiring data from a diverse set of matrices:
human whole blood, plasma, serum, urine, fecal water, CSF and tissue extracts
murine plasma, serum, urine, brain homogenates and tissue extracts
porcine and bovine plasma
mammalian cell, bacterial and yeast extracts
insect extracts
macro and microalgal extracts
Sample are preferably delivered either in individual cryo vials (10.8 or 12 mm diameter) or in a deepwell plate format to enable facile use of lab automation for NMR sample preparation.
For urine, a volume of 600µl or more is preferred, but 200µl is acceptable.
For plasma or serum, a volume of 325µl is preferred but 110 µl is acceptable.
For extracts or other matrices, volumes of 100-400 µl is needed, depending on spectrometer and project-specific limitations.
All projects are discussed before commitment, ventilating strategy, sample requirements, costs etc.
Currently, there are automated protocols for methanol extraction of plasma/serum, BUME lipid extraction and Folch lipid extraction. All protocols are set up on an Agilent Bravo 96-channel platform ("Boba") so works in a 96-well plate or tube format.
All projects need to cover consumables cost, currently the fee is 220 SEK/sample.
For service projects, 10 h of free staff time is included. This usually covers sample preparation, data acquisition, compilation and delivery. Any staff time in excess of 10 h is charged with 700 SEK/h.
Collaborative projects have 40 h staff time included, otherwise policy is the same as for service projects.
Data acquisition relies heavily on automation, from sample handling to actual spectrometer operation, to be able to run 24/7 without human intervention. Spectrometers at SNC used for metabolomics are equipped with cooled SampleJets and can hold up to 480 short NMR samples. In practice, sample preparation and submission for data acquisition is done in a pace suited to match the spectrometer throughput.
For the vast majority of projects, acquisition means running different 1-dimensional 1H-NMR experiments on each sample. Typically 1D-NOESY (metabolites and proteins/high molecular weight complexes detected), 1D-CPMG (only metabolites detected) and 1D diffusion-edited (only high molecular weight components detected) experiments are acquired for each sample depending on the particular matrix.
Frequently asked questions: data acquision
For Bruker IVDr plasma/serum projects, throughput is approximately 45 sampels/day. For other matrices and use of the 800 MHz 'indigo' spectrometer, throughput is a bit higher, up to 70 samples/day.
For profiling, typically 1D NOESY and CPMG-edited 1H experiments are used. Under the Bruker IVDr SOP on plasma or serum, also 2D J-resolved and 1D diffusion-edited experiments JEDI and DIRE are acquired.
For annotation purposes, 2D natural abundance 1H,13C-HSQC, 1H,13C-HMBC and 2D 1H,1H-TOCSY are frequently employed on selected samples of a given project to aid in identification of metabolites. 2D data is acquired on the high-field cryoprobe-equipped spectrometers at SNC GU.
3. Spectral processing
The acquired data is transformed from time to frequency domain via a Fourier transform often followed by baseline correction and smoothing. In the spectra different metabolites have characteristic peak patterns with intensities proportional to their corresponding concentrations, in theory making both annotations and concentration quantification straight forward. Often peaks from different metabolites overlap however, which in combination with slight peak position changes between spectra due to pH and ionic strength differences make this more difficult requiring either deconvolution to separate signals from different metabolites, or a less ambitious approach dividing the spectrum into buckets accepting that some measured values are sums of concentrations from different metabolites.
Manual spectral deconvolution
An example of manual spectral deconvolution and metabolite identification in a serum sample, using software and spectral reference databases. It's clear that overlapping signals can be a significant problem in complex biofluids, e.g. the three signal clusters for glycerol (in green) overlapping with glucose and other metabolite signals.
Automated spectral deconvolution
An example of automated spectral deconvolution using the TopSpin function mldcon (Bruker BioSpin) on cerebrospinal fluid.
4. Data analysis and metabolite identification
Typically, first the overall quality and general trends in the data are determined. This is often assessed with unsupervised multivariate analysis, like principal component analysis (PCA).
Case study: principal component analysis
Already from a PCA (A, B) score plot it is evident that medical therapy alone (MTA) and combined gastric bypass plus medical therapy (CSM) at six months differ from each other and baseline. (C) From the corresponding loading plot some molecules important for the separation can be found.
Second, any associations between the data and independent variables are often described using supervised multivariate analysis like Projections to latent structures (PLS), Orthogonal Projections to latent structures (OPLS), or Random Forest.
Case study: supervised modeling
A PLS model of the baseline vs CSM month 6 samples showing a convincing (A) receiver operating characteristic curve, (B) confusion matrix, and (D) permutation model all supporting that there is a difference between the groups. (C) Using between minimal-optimal and all relevant number of variables give the best separation with optimal validation performance. (E) The most important variables (metabolites) for the model and if they are higher or lower in concentrations six months after surgery.
Third, traditional univariate analysis looking for individual metabolite associations to independent variables is done.
Case study: univariate statistics
(A-D) Univariate comparisons between baseline, and MTA and CSM at six months. (C) Univariate analysis to a large extent often gives similar results as multivariate analysis.
Frequently asked questions: data analysis
When peaks from different metabolites don’t overlap and the concentrations are sufficient (~10 µM and above), any error should be in the range of some percent. When peaks are overlapping the error is larger and also much more difficult to estimate. In the particular case of analysing serum, plasma, or urine samples using the Bruker IVDr protocol, error estimates are available.
Often all peaks (assigned and unassigned) are treated the same in the analysis. Having unassigned peaks means that molecules with many peaks can get larger weight in multivariate analysis. This is often neglected or taken care of using variable reduction techniques.
The intensity to concentration factor requires some work to obtain and therefore often the intensities, ‘relative concentrations’, are used as is. In urine, but also many other matrices, normalization between biological samples is often more important due to different overall concentrations.
A combination can improve the understanding of the data. Univariate analysis provides a basic overview, while multivariate analysis delves deeper into complex relationships and interactions among variables.
5. Biological interpretation
The biological interpretation of the metabolomics output and analysis is generally left to the client but SNC is happy to refer to the relevant bioinformatic competence at NBIS.
Case study
NMR urine metabolomics implicate gastric bypass‐induced urinary metabolomic changes in host–microbial co‐metabolism, aromatic amino acid metabolism, and BCAA catabolism in the enhanced improvements in blood pressure, glycaemia, dyslipidaemia, and albuminuria observed in patients with obesity, type 2 diabetes, and early chronic kidney disease treated with medical therapy.
The instrumentation at the site in Gothenburg includes cooled sample changers hooked up to 600MHz and 800 MHz spectrometers. The 600MHz spectrometer is set up as a Bruker In Vitro Diagnostics for research (IVDr) platform (more information from Bruker) making results obtained in Gothenburg directly comparable to results obtained elsewhere in the world, allowing for big collaborations as well as longitudinal studies where samples are recorded at different times.
The workflow guarantess minimal degradation and high quality of samples using highly automated procedures with an unbroken cold chain, generating NMR data of high international standard which is essential for downstream biological interpretation of metabolomics data.
When running plasma/serum samples under IVDr SOP, every sample is analyzed for spectral quality as well as sample quality. Apart from checking that the NMR-spectroscopy is performed according to the IVDr specifications, the identity and integrity of the actual sample is also assessed. Preanalytical mishandling of the sample or even contamination from the actual blood draw can be detected and that can in turn be taken into consideration when analyzing the metabolite and lipoprotein profiling data.
The reproducibility of metabolomics measurement by adhering to the IVDr standard operating procedures is excellent, as shown here for spectra of a quantification reference between Apr 2018 and Mar 2020 where the CV is well below 5%.
We offer access to metabolomics measurements in projects ranging from pure service to full statistical analysis of data.
For metabolomics, a per-sample cost model is employed. The latest price revision is effective from Jan 1st 2026. The current rate is 350 SEK/sample.
Depending on the agreed support level at project start, additional costs for staff hours may apply. For pure service projects (sample handling, data acquisition, data processing and data delivery) 10 hours of free staff time is included with the requirement that the NMR Centre is acknowledged in any resulting publication using the data. For collaborative projects 40 hours of free staff time is included with the requirement that at least one NMR Centre staff is included as coauthor on any resulting publication. In both cases, staff time spent in excess of the free 10 and 40 hours, respectively, will be charged at 700 SEK per hour.
Plasma/serum, sample preparation through dilution 50/50:
B.I.Lisa automatic lipoprotein profiling (IVDr)
B.I.Quant automatic quantitation of ~40 metabolites (IVDr)
Biobank QC quality control of sample matrix and data quality (IVDr)
At least 110 µl sample, preferably 325 µl
Plasma/serum, sample preparation through methanol precipitation:
Automatic deproteinization with methanol precipitation using an Agilent Bravo liquid handling robot
Compatible with internal standard quantitation
Offers additional metabolites compared to dilution sample preparation
At least 150 µl sample
Urine, sample preparation through dilution 10/90:
B.I.Quant automatic quantitation of up to 150 metabolites (IVDr)
Biobank QC quality control of sample matrix and data quality (IVDr)