A new force measurement device for evaluating finger extension function in the healthy and rheumatoid arthritic hand

Artikel i vetenskaplig tidskrift
Författare Sofia Brorsson
A. Nilsdotter
Christer Sollerman
A. J. Baerveldt
M. Hilliges
Publicerad i Technology and Health Care
Volym 16
Nummer/häfte 4
Sidor 283-92
ISSN 0928-7329
Publiceringsår 2008
Publicerad vid
Sidor 283-92
Språk en
Ämneskategorier Handkirurgi

Sammanfattning

Although often neglected, finger extension force is of great importance for developing grip strength. This paper describes the design and evaluation of a new finger extension force measurement device (EX-it) based on the biomechanics of the hand. Measurement accuracy and test-retest reliability were analysed. The device allows measurements on single fingers as well as all the fingers (excluding the thumb) of both healthy and deformed hands. The coefficient of variation in the device was 1.8% of the applied load, and the test-retest reliability showed a coefficient of variation no more than 7.1% for healthy subjects. This study also provides reference values for finger extension force in healthy subjects and patients with rheumatoid arthritis (RA). Significant differences were found in extension strength between healthy subject and RA patients (men, p < 0.05 and women, p < 0.001). EX-it provides objective and reliable data on the extension force capacity of normal and dysfunctional hands and can be used to evaluate the outcome of therapeutic interventions after hand trauma or disease.

Sidansvarig: Webbredaktion|Sidan uppdaterades: 2012-09-11
Dela:

På Göteborgs universitet använder vi kakor (cookies) för att webbplatsen ska fungera på ett bra sätt för dig. Genom att surfa vidare godkänner du att vi använder kakor.  Vad är kakor?